General
AIM-M100 medical intelligent insulation monitoring instrument adopts advanced microcontroller technology, which has high integration, compact size, convenient installation and integrates intelligence, digitalization and networking in one. It is ideal selection for insulation monitoring of isolation power system in Class 2 medical.
Product Functions
Ø Functions of real-time monitoring and fault alarming of the ground insulation resistance, transformer load current and transformer winding temperature of the monitored IT system;
Ø Can be used with insulation fault locator, remotely starting fault-locating and displaying locating results when there are insulation faults;
Ø Real-time monitor the line disconnection fault, temperature sensor disconnection fault and the functional grounding line disconnection fault of the monitored system, and give the alarm indication when the fault occurs;
Ø Relay alarm output, LED alarm indication and other faults indication functions;
Ø Two kinds of fieldbus communication technology, which are used for centralized alarm and display instrument, test signal generator, insulation fault locator and upper computer management software communications, and can monitor the operation status of IT system in real time.
Ø Events logging function, which can record alarm occurrence time and fault type and is convenient for operation personnel to analyze the operation conditions of system and promptly eliminate the faults;
Technical Parameter
Technical parameters of AIM-M100 insulation monitoring instrument
Auxiliary power supply |
Voltage |
AC220V (fluctuating range±10%) |
Temperature measurement |
Thermistor |
PT100 |
Frequency |
50/60Hz |
Measuring range |
-50—+200℃ |
||
Maximum power consumption |
<8W |
Alarm value scope |
0—+200℃ |
||
Insulation monitoring |
Measuring range of insulation resistance |
10-999kΩ |
Alarm output |
Output mode |
2-route relay output(programmable) |
Absolute percentage error |
0—±10% |
Contact capacity |
AC 250V/3A DC 30V/3A |
||
Alarm value scope |
50—999kΩ |
Environment |
Operating temperature |
-10—+55℃ |
|
Response time |
<2s |
Storage temperature |
-20—+70℃ |
||
Measuring voltage |
<12V |
Relative humidity |
5%-95%, non-condensate |
||
Measuring current |
<50uA |
Altitude |
≤2500m |
||
Load current |
Measuring range |
2.1-50A |
Communication |
RS485 interface, Modbus-RTU agreement |
|
Alarm value scope |
5-50A |
Rated impulse voltage/pollution degree |
4KV/Ⅲ |
||
|
Measuring accuracy |
≤±5% |
EMC electromagnetic compatibility/electromagnetic radiation |
Conform to IEC 61326-2-4 |
Wiring
Dimension drawings
Product selection
Type |
Picture |
Description |
|
AITR series medical isolation transformer |
|
AITR series isolation transformer is specially used in medical IT system. The windings are treated with double insulation and have electrostatic shielding layer, which reduces electromagnetic interference between windings. The PT100 temperature sensor is installed in the wire bag to monitor the temperature of transformer. The whole body is treated with vacuum invasion paint, which increases mechanical strength and corrosion resistance. The product has good temperature rise performance and very low noise. |
|
AIM-M100 medical intelligent insulation monitoring instrument |
The AIM-M100 medical intelligent insulation monitor is compact in size, easy to install, intelligent, digital and networked, and is an ideal choice for insulation monitoring of isolation power supply systems in operating rooms, intensive care units and other medical places. |
||
AKH-0.66P26 current transformer |
|
The AKH-0.66P26 type current transformer is the protective current transformer supporting the AIM-M100 insulation monitor, of which the maximum measurable current is 60A and the transformation ratio is 2000:1. The current transformer is directly fixed inside cabinet by screwing, and the secondary side is leaded out by the terminal, which is convenient to install and use. |
|
ACLP10-24 dc power module |
|
Special DC module for instrument, stable output voltage. The module adopts the standard guide way to install, and can be installed on the same guide way with the insulation monitor, easy to install. |
|
AID series external alarm and display instrument
|
AID120 |
|
It is suitable for wall installation embedded in operating room or nurse station and can monitor 1 AIM-M100 insulation monitor. It has sound and light alarm function of insulation, overload, overtemperature and equipment failure, digital tube display and RS485 communication |
AID150 |
|
LCD display, RS485 bus, centralized monitoring of up to 16 SETS of AIM-M100 medical intelligent insulation monitor data, sound and light alarm can be remote. The AID150 can also monitor data from multiple AIM-R100 residual current monitors. |